Automated balloon control in resuscitative endovascular balloon occlusion of the aorta (REBOA)

Loading...
Thumbnail Image
Files
8367.pdf(873.29 KB)
Accepted version
Date
2018-11-01
Authors
McCarthy, Cormac
Kanterman, Ian
Trauzettel, Fabian
Jaeger, Herman Alexander
Goetz, Alice-Anne
Colvard, Benjamin
Swanstrom, Lee
Cantillon-Murphy, Pádraig
Journal Title
Journal ISSN
Volume Title
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Research Projects
Organizational Units
Journal Issue
Abstract
Objective: The goal of this work was to demonstrate technical feasibility of automated balloon pressure management during REBOA in the pre-clinical setting. Methods: This paper presents an intelligent balloon management device which automates the balloon inflation process, preventing the possibility of balloon over or under inflation, optimizes inflation pressure and if indicated, deflates automating partial REBOA to allow distal organ perfusion. Edwards TruWave pressure transducers are used to monitor the blood pressure proximal and distal to the balloon, as well as the internal balloon pressure. A faux PID controller, implemented on an Arduino platform, is used in a feedback control loop to allow a user defined mean arterial pressure setpoint to be reached, via a syringe driver which allows intelligent inflation and deflation of the catheter balloon. Results: Ex vivo testing on a vascular perfusion simulator provided the characteristic behavior of a fully occluded aorta, namely the decrease of distal pressure to zero. In vivo testing on live porcine models indicated that automated partial REBOA is achievable and by enabling partial occlusion may offer improved medical outcomes compared to manual control. Conclusion: Automated balloon pressure management of endovascular occlusion is feasible and can be successfully implemented without changes on current clinical workflows. Significance: With further development, automated balloon management may significantly improve clinical outcomes in REBOA.
Description
Keywords
Blood pressure , Biomedical monitoring , Testing , Catheters , Arteries , Sensors , Monitoring , REBOA , Balloon catheter , Inflation , Partial REBOA , Feedback , Automation , Syringe , Pressure
Citation
McCarthy, C., Kanterman, I., Trauzettel, F., Jaeger, H. A., Goetz, A., Colvard, B., Swanstrom, L. and Cantillon-Murphy, P. (2018) 'Automated Balloon Control in Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA)', IEEE Transactions on Biomedical Engineering, In Press, doi: 10.1109/TBME.2018.2878642
Copyright
© 2018 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.