Estimation of the number of scans required per hard-to-clean location and establishing the limit of quantification of a partial least squares calibration model when the FTIR is used for pharmaceutical cleaning verification

Loading...
Thumbnail Image
Files
molecules-27-04569_(1).pdf(1.06 MB)
Published Version
Date
2022-07
Authors
Sarwar, Apu
McSweeney, Conor
Smith, Mark
Moore, Eric
Journal Title
Journal ISSN
Volume Title
Publisher
MDPI
Research Projects
Organizational Units
Journal Issue
Abstract
This study aims to identify two critical components required for pharmaceutical cleaning verification when an FTIR is used: (a) the number of scans required per hard-to-clean location, and (b) the limit of quantification (LOQ) of the FTIR instrument when measuring the surface contamination. The current practice in pharmaceutical manufacturing does not require multiple samples as it is standard practice to collect a single swab sample from a 25 x 25 cm area from a difficult-to-reach area of the manufacturing equipment. However, since the FTIR will only scan a tiny portion of the surface compared to the swab, a sufficient number of samples (data points) are required to provide enough confidence to ensure that the measurement results are close to the true value with a maximum degree of certainty. Similarly, calculating the LOQ for a linear regression could be straightforward. However, complexity arises when the experimental data are complex; in this case, the complexity arises due to the nature of the measurement and the lack of the defined peak in the pre-processed spectra. Therefore, this study uses the practical approach of calculating the sample size and the LOQ.
Description
Keywords
Reflection-absorption spectroscopy , Varying roughness , Trace analysis , Surfaces
Citation
Sarwar, A., McSweeney, C., Smith, M. and Moore, E. (2022) ‘Estimation of the number of scans required per hard-to-clean location and establishing the limit of quantification of a partial least squares calibration model when the FTIR is used for pharmaceutical cleaning verification’, Molecules, 27(14), 4569 (9pp). doi: 10.3390/molecules27144569
Link to publisher’s version