A deep neural network for 12-lead electrocardiogram interpretation outperforms a conventional algorithm, and its physician overread, in the diagnosis of atrial fibrillation
Smith, Stephen W.; Rapin, Jeremy; Li, Jia; Fleureau, Yann; Fennell, William; Walsh, Brooks M.; Rosier, Arnaud; Fiorina, Laurent; Gardella, Christophe
Date:
2019-09-08
Copyright:
©2019 The Authors. Published by Elsevier B.V. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).Contents lists available atScienceDirectIJC Heart & Vasculaturejournal homepage:http://www.journals.elsevier.com/ijc-heart-and-vasculature
Citation:
Smith, S. W., Rapin, J., Li, J., Fleureau, Y., Fennell, W., Walsh, B. M., Rosier, A., Fiorina, L. and Gardella, C. (2019) 'A deep neural network for 12-lead electrocardiogram interpretation outperforms a conventional algorithm, and its physician overread, in the diagnosis of atrial fibrillation', IJC Heart & Vasculature, 25,100423 (6pp.). DOI: 10.1016/j.ijcha.2019.100423
Abstract:
Background: Automated electrocardiogram (ECG) interpretations may be erroneous, and lead to erroneous overreads, including for atrial fibrillation (AF). We compared the accuracy of the first version of a new deep neural network 12-Lead ECG algorithm (Cardiologs®) to the conventional Veritas algorithm in interpretation of AF. Methods: 24,123 consecutive 12-lead ECGs recorded over 6 months were interpreted by 1) the Veritas® algorithm, 2) physicians who overread Veritas® (Veritas® + physician), and 3) Cardiologs® algorithm. We randomly selected 500 out of 858 ECGs with a diagnosis of AF according to either algorithm, then compared the algorithms' interpretations, and Veritas® + physician, with expert interpretation. To assess sensitivity for AF, we analyzed a separate database of 1473 randomly selected ECGs interpreted by both algorithms and by blinded experts. Results: Among the 500 ECGs selected, 399 had a final classification of AF; 101 (20.2%) had ≥1 false positive automated interpretation. Accuracy of Cardiologs® (91.2%; CI: 82.4–94.4) was higher than Veritas® (80.2%; CI: 76.5–83.5) (p < 0.0001), and equal to Veritas® + physician (90.0%, CI:87.1–92.3) (p = 0.12). When Veritas® was incorrect, accuracy of Veritas® + physician was only 62% (CI 52–71); among those ECGs, Cardiologs® accuracy was 90% (CI: 82–94; p < 0.0001). The second database had 39 AF cases; sensitivity was 92% vs. 87% (p = 0.46) and specificity was 99.5% vs. 98.7% (p = 0.03) for Cardiologs® and Veritas® respectively. Conclusion: Cardiologs® 12-lead ECG algorithm improves the interpretation of atrial fibrillation.
Show full item record